The
purpose of the Steam Dryness Value Test is to ensure that
an acceptable amount of moisture is present in the steam
supply. If too little moisture were present, superheating
of the steam may occur. Too little moisture may prevent
sterilization conditions within the load, particularly because
moisture is important in breaking down the cell structure
of sporing organisms. If too much moisture is present, the
load may not be able to be dried by completion of the cycle.
Under the correct conditions wrapped loads are able to be
stored for considerable periods in a sterile state. This
is only true however where the wrapping material maintains
its integrity by providing a microbiological barrier. If
the load and its wrapping are damp when removed from the
sterilizer this barrier is broken and can provide a path
for re-contamination.
The method employed is the heat balance using a stainless
steel vacuum flask.
The principal is that the flask is primed with a known mass
of water at a known temperature. Steam is condensed in the
flask thus raising the water temperature. The final mass
and temperature of the water are then measured and placed
into a calculation. If the final water temperature was lower
or the final mass was greater, the steam would be 'wetter'
(having a lower value).
Whilst the method is not regarded as a truly accurate measurement
of moisture in the steam, it can be used to demonstrate
acceptable dryness for sterilization purposes. A dryness
value of 1.0 is equivalent to Dry Saturated Steam. For sterilization
purposes, a factor of 0.9 is acceptable for nonmetal sterilizer
loads. For loads containing metal, a factor of 0.95 is considered
to be acceptable.
As the flask has to be connected within 0.5m of the sampling
point, SVMS have incorporated a flask holder in the kit
so that the flask can be mounted in a convenient position.
This holder also serves as a measuring container to prime
the flask.
For the engineers' convenience, the calculation can be performed
manually or by the use of a computer spreadsheet. An electronic
scientific calculator and CD containing all forms and calculations
are supplied.
•